The MR64 regulator is specially designed for use with medical oxygen breakdown panels. The specified flow capacity is based on the medical oxygen pipeline performance requirements as details in AS2896, ie. the maximum flow achievable at an outlet pressure of 375 kPa when starting from a static outlet pressure of 450 kPa.
Specifications:
Max. Inlet Pressure: | 3,000 kPa @ 15˚C |
Outlet Pressures: | 415 kPa nominal |
Outlet Gauge: | 600 or 800 kPa (end users preference) 63mm diameter brass, STST case |
Inlet Fitting: | 1” BSP nut/nipple (o-ring seal) |
Outlet Fitting: | 1/2”NPT female, (other fittings on request) |
Weight: | 2.7 kg |
Materials:
Body: | Chrome plated brass bar stock |
Bonnet: | Chrome plated brass bar stock |
Seat: | Viton |
Diaphragm: | Neoprene |
Inlet Filters: | Bronze 63μm |
O-ring seals: | Viton |
Reviews
There are no reviews yet.